Softening Point Tester of Bitumen
Model STL 206
Softening Point Tester of Bitumen is a ring-and-ball apparatus for determining bitumen softening point.
It covers the range from 30 to 150°C for petroleum products.
The magnetic stirrer and heater has separate regulators for stirrer speed and temperature rate.
It includes 4 brass rings, 4 steel balls, and a ring-holder for 2 rings.
Two thermometers cover the ranges from -2 to +80°C and 30 to 200°C.
The Softening Point Tester of Bitumen determines the softening point of bitumen by the ring-and-ball method. It covers the range from 30 to 150°C for petroleum products.
Product Overview
The Softening Point Tester of Bitumen is used for determining the softening point of bitumen in the range from 30 to 150°C of petroleum products. It uses a magnetic stirrer and heater with two regulators, one for stirrer speed and one for the rate of temperature change, set at 5 ± 0.5°C increase. The kit includes brass square-shouldered rings, steel balls with brass centering guides, a ring-holder, a pouring plate, a bath vessel, glass beakers with stands or clamps, and two thermometers.
Softening Point Tester of Bitumen STL 206: Technical Specifications
| Parameter | Value |
|---|---|
| Test Method | Ring-and-ball apparatus |
| Softening Point Range | 30 to 150°C |
| Heater Power Supply | 220 V AC |
| Heating Rate | 5 ± 0.5°C increase |
| Stirrer Control | Separate regulator for speed |
| Temperature Rate Control | Separate regulator for rate of change |
| Ring Material | Brass |
| Ring Type | Square-shouldered, 4 nos |
| Ring Top Inside Diameter | 19.8 ± 0.3 mm |
| Ring Bottom Inside Diameter | 15.9 ± 0.3 mm |
| Ring Total Depth | 6.4 ± 0.4 mm |
| Ball Material | Steel, 4 nos |
| Ball Diameter | 9.5 mm |
| Ball Weight | 3.5 ± 0.05 g |
| Ball Centering Guide | Brass, 4 nos |
| Ring-Holder Material | Brass, stainless steel, or aluminium |
| Ring-Holder Capacity | 2 rings |
| Ring to Bottom Plate Gap | 25.4 ± 0.8 mm |
| Bottom Plate to Beaker Distance | 12.7 to 19 mm |
| Pouring Plate | Flat brass, approx. 50 × 75 mm |
| Bath Vessel Depth | Not less than 120 mm |
| Bath Heating Rate | 5°C/min |
| Bath Variation Limit | ± 0.5°C per 1-min period after first 3 minutes |
| Beaker Quantity | 3 nos |
| Beaker Diameter | Not less than 85 mm |
| Beaker Depth | Not less than 120 mm |
| Beaker Fill Depth | 100 to 108 mm |
| Low Range Thermometer | -2 to +80°C, sensitivity 0.5°C, 1 no |
| High Range Thermometer | 30 to 200°C, sensitivity 1°C, 1 no |
Technical Data
- Heating and Stirring
- Magnetic stirrer and heater: 220 V AC
- Regulator 1: stirrer speed
- Regulator 2: rate of temperature change, 5 ± 0.5°C increase
- Rings
- Material: brass
- Quantity: 4 nos, square-shouldered
- Top inside diameter: 19.8 ± 0.3 mm
- Bottom inside diameter: 15.9 ± 0.3 mm
- Total depth: 6.4 ± 0.4 mm
- Balls and Guides
- Balls: steel, 4 nos
- Ball diameter: 9.5 mm
- Ball weight: 3.5 ± 0.05 g
- Centering guides: brass, 4 nos
- Ring-Holder
- Material: brass, stainless steel, or aluminium
- Capacity: 2 rings
- Gap between ring bottom and bottom plate: 25.4 ± 0.8 mm
- Distance from bottom plate to beaker: 12.7 to 19 mm
- Thermometer bulb level with ring bottom and within 12.7 mm of rings
- Immersion depth adjustable with screw/nut
- Pouring Plate and Bath
- Pouring plate: flat, smooth brass, approx. 50 × 75 mm
- Bath vessel: glass, heatable, not less than 120 mm deep
- Bath heating rate: 5°C/min
- Permissible variation: ± 0.5°C per 1-min period after first 3 minutes
- Beakers
- Quantity: 3 nos
- Diameter: not less than 85 mm
- Depth: not less than 120 mm
- Fill depth: 100 to 108 mm
- Stand/clamp included for setting thermometer
- Thermometers
- Low softening point: -2 to +80°C, sensitivity 0.5°C, 1 no
- High softening point: 30 to 200°C, sensitivity 1°C, 1 no
Key Features
- Softening Point Range: 30 to 150°C
- Test Method: ring-and-ball apparatus
- Heating Rate Control: 5 ± 0.5°C increase with separate regulator
- Stirrer Speed: separate magnetic stirrer regulator
- Ring Set: 4 brass square-shouldered rings
- Ball Set: 4 steel balls, 9.5 mm diameter
- Thermometers: low range (-2 to +80°C) and high range (30 to 200°C)
Experiments
Softening point determination of bitumen by the ring-and-ball method, from 30 to 150°C.
Construction and Design
The Softening Point Tester of Bitumen uses brass rings and a brass or stainless steel or aluminium ring-holder. The magnetic stirrer and heater has two regulators. The ring-holder allows adjustment of thermometer position and immersion depth.
Scope of Delivery
- Magnetic stirrer and heater with two regulators, 220 V AC
- Brass square-shouldered rings, 4 nos
- Steel balls, 4 nos, 9.5 mm diameter, 3.5 ± 0.05 g
- Brass ball centering guides, 4 nos
- Ring-holder with provision for 2 rings
- Flat brass pouring plate, approx. 50 × 75 mm
- Glass bath vessel, not less than 120 mm deep
- Glass beakers, 3 nos, with stand/clamp
- Low softening point thermometer, -2 to +80°C, sensitivity 0.5°C, 1 no
- High softening point thermometer, 30 to 200°C, sensitivity 1°C, 1 no
What is the softening point range of this tester?
It covers 30 to 150°C for petroleum products.
What test method does it use?
It uses the ring-and-ball apparatus.
How many rings and balls are included?
It includes 4 brass rings and 4 steel balls.
What thermometers are supplied?
One low softening point thermometer (-2 to +80°C, 0.5°C sensitivity) and one high softening point thermometer (30 to 200°C, 1°C sensitivity).
What is the heating rate?
The bath heating rate is 5°C per minute.
Need Assistance? We're Here to Help!
Connect with our team for product specifications, pricing, and customized solutions for your institution or project.