🔎 ISO 9001:2015 & CE Certified Lab Equipment Manufacturer — Get CIF Quote in 24 Hours → WhatsApp Now
ISO 9001:2015 Certified CE Marked Equipment Exporting Since 1993 60+ Countries Served Quote in 24 Hours

Series and Parallel Pumps Experiment: Procedure, Characteristics and Results

Aim of the Experiment

  • To determine the performance characteristics (Head vs Discharge, Q vs Efficiency) of centrifugal pumps operating individually, in series, and in parallel
  • To compare the effect of series and parallel configurations on total head and flow rate

Theory

When pumps are connected in series or parallel, the combined performance differs from that of a single pump:

Pumps in Series

The total head increases while the flow rate stays the same as a single pump. The combined H-Q curve is obtained by adding the head values of each pump at the same flow rate:

Hseries = H₁ + H₂ (at same Q)

Series configuration is used when high head is required at moderate flow — e.g., booster pumps in tall buildings.

Pumps in Parallel

The total flow rate increases while the head stays the same. The combined Q-H curve is obtained by adding flow rates at the same head:

Qparallel = Q₁ + Q₂ (at same H)

Need lab equipment for your institution?
ISO 9001:2015 & CE certified manufacturer — exported to 60+ countries since 1993. Get pricing, datasheet & a CIF quote.
✓ Reply within 4 business hours · ✓ No obligation · ✓ Free quotation

Parallel configuration is used when high flow rate is required at the same head — e.g., water distribution networks.

Apparatus

  • Multi-pump test rig (FluidoSurge-X-192) with two identical centrifugal pumps
  • Manifold piping system with valves for series/parallel switching
  • Pressure gauges at pump inlet and outlet
  • Flow meter (rotameter or electromagnetic)
  • Wattmeter or power analyser for shaft power measurement
  • Control valve on delivery line for varying flow resistance

Procedure

Single Pump Characteristics

  1. Configure the valve arrangement for single pump operation. Keep Pump 1 running; isolate Pump 2.
  2. Fully open the delivery valve. Record flow rate (Q), inlet pressure (Pi), outlet pressure (Po), and power input (Pw).
  3. Gradually close the delivery valve in steps to reduce flow rate. At each step, record Q, Pi, Po, and Pw.
  4. Continue until the delivery valve is fully closed (shut-off head condition, Q = 0).
  5. Calculate head at each point: H = (Po − Pi) / ρg + velocity head difference.

Pumps in Series

  1. Reconfigure the valve manifold to connect both pumps in series (outlet of Pump 1 feeds inlet of Pump 2).
  2. Start both pumps. Repeat the same procedure — vary delivery valve position and record readings at 6–8 flow rates.

Pumps in Parallel

  1. Reconfigure the valve manifold for parallel connection (both pump outlets join a common delivery header).
  2. Start both pumps. Record Q, pressure, and power at each delivery valve position as before.

Observation Table

Config Q (L/min) Pi (bar) Po (bar) H (m) Pw (W) η (%)
Single
Series
Parallel

Calculations

Total Head: H = (Po − Pi) × 10 / ρg (m of water, when pressures in bar)

Hydraulic Power Output: Ph = ρgQH (W)

Overall Efficiency: η = Ph / Pw × 100%

Graph

Plot three H-Q curves on the same graph (Single, Series, Parallel). Observations:

  • Series curve: same shape as single pump curve but shifted vertically upward (doubled head at each Q)
  • Parallel curve: same shape but shifted horizontally to the right (doubled Q at each H)

Precautions

  • Never run centrifugal pumps dry — ensure the system is primed before starting.
  • Open the delivery valve before starting pumps in series to avoid pressure build-up.
  • Change flow rates slowly and allow the system to stabilise before recording readings.
  • Check that both pumps rotate in the correct direction.
Need this apparatus for your lab?

Scientico India manufactures this and the full range of Fluid Mechanics Lab equipment — CE-certified and supplied to engineering colleges worldwide.

Explore Fluid Mechanics Lab Equipment → or request a quote

  • Impact of Jet on Vanes Experiment — Apparatus and Force Calculations
  • Leave a Comment

    Your email address will not be published. Required fields are marked *

    Make An Enquiry